TWiki> Public Web>KnownIssues>KnownIssuePcap01 (revision 1)EditAttach

Fixing problem with monitoring restart

ID: KBPCAP200708

Analyzer type: SNT-7531, SNTlite, SNTsmart

Functional area: Ethernet monitoring option ()

Problem description: There are no monitoring packets arrived after ethernet monitoring restart, filtering disabled.

Possible cause: WinPCAP packet version prior to 4.0.1 installed on a PC, you need to update it. To check this, press Start -> Settings -> Control Panel and start "Add/Remove programs" applet. Examine installed program listing and search for WinPCAP and its version.

Solution: Download 4.0.1 version form winpcap.org (or from version archive), shutdown SNT software, then start downloaded packet installation, following installation program instructions. After succesefull WinPCAP instllation, start SNT software again.

EditorForm
ArticleStatus? Editing
TranslationStatus? Editing
Comment

Edit | Attach | Print version | History: r2 < r1 | Backlinks | Raw View | More topic actions...
Topic revision: r1 - 2007-08-10 - 16:29:52 - IlyaDanilov
 
© ООО "НТЦ Севентест". 2007. Перепечатка и копирование материалов без письменного разрешения правообладателя запрещены.
Есть идеи, вопросы или проблемы, связанные с TWiki? Свяжитесь с нами.